- Field Metrology Well, 33 °C to 350 °C, w/9143-INSX, w/Process Electronics
- Temperature Range at 23 °C: 33 °C to 350 °C (91 °F to 662 °F)
- Display Accuracy: ± 0.2 °C Full Range
- Stability: ± 0.02 °C at 33 °C ± 0.02 °C at 200 °C ± 0.03 °C at 350 °C
- Axial Uniformity at 40 mm (1.6 in): ± 0.04 °C at 33 °C ± 0.1 °C at 200 °C ± 0.2 °C at 350 °C
- Radial Uniformity: ± 0.01 °C at 33 °C ± 0.015 °C at 200 °C ± 0.02 °C at 350 °C
- Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes): ± 0.015 °C Full Range
- Hysteresis: 0.03°C
- Operating Conditions: 0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
- Environmental Conditions (for all specifications except temperature range): 13 °C to 33 °C
- Immersion (Well) Depth: 150 mm (5.9 in)
- Insert OD: 25.3mm (1.00 in)
- Heating Time: 5 min: 33 °C to 350 °C
- Cooling Time: 32 min: 350 °C to 33 °C 14 min: 350 °C to 100 °C
- Resolution: 0.01 °
- Display: LCD, °C or °F user-selectable
- Size (H x W x D): 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
- Weight: 7.3 kg (16 lb)
- The process version provides a built-in two-channel thermometer readout that measures PRTs, RTDs, thermocouples, and 4-20 mA transmitters which includes the 24 V loop supply to power the transmitter.
>
Each Unit Includes
- Insert A
- RS-232 and 9930 Interface-it control software
- Test Leads